| Company Announcement - Philips is initiating a voluntary recall notification* to ensure patient safety in consultation with regulatory agencies
- Corrective actions include the deployment of updated instructions for use and a repair and replacement program for affected devices
- Philips aims to address all affected devices within the scope of this correction as expeditiously as possible
Amsterdam, the Netherlands– Following the company update on April 26, 2021, Royal Philips (NYSE: PHG; AEX: PHIA) today provides an update on the recall notification* for specific Philips Bi-Level Positive Airway Pressure (Bi-Level PAP), Continuous Positive Airway Pressure (CPAP), and mechanical ventilator devices to address identified potential health risks related to the polyester-based polyurethane (PE-PUR) sound abatement foam component in these devices. The majority of the affected devices within the advised 5-year service life are in the first-generation DreamStation product family. To date, Philips has produced millions of Bi-Level PAP, CPAP and mechanical ventilator devices using the PE-PUR sound abatement foam. Despite a low complaint rate (0.03% in 2020), Philips determined based on testing that there are possible risks to users related to this type of foam. The risks include that the PE-PUR foam may degrade into particles which may enter the device's air pathway and be ingested or inhaled by the user, and the foam may off-gas certain chemicals. The foam degradation may be exacerbated by use of unapproved cleaning methods, such as ozone,** and high heat and high humidity environments may also contribute to foam degradation. | | | |